- High accuracy at low ranges (below 0.05 NTU)
- Two, three or four point calibration
- USB and RS232 PC connectivity
- Log up to 200 records
- GLP Features
- User friendly, backlit display with guidance codes
- Battery % on startup
- Continuous current time on display
Specifications:
- Range 0.00 to 9.99; 10.0 to 99.9 and 100 to 1000 NTU
- Range Selection automatic
- Resolution 0.01 NTU from 0.00 to 9.99 NTU; 0.1 NTU from 10.0 to 99.9 NTU; 1 NTU from 100 to 1000 NTU
- Accuracy @25°C ±2% of reading plus 0.02 NTU
- Repeatability ±1% of reading or 0.02 NTU, whichever is greater
- Stray Light < 0.02 NTU
- Light Detector silicon photocell
- Light Source tungsten filament lamp
- Lamp Life greater than 100,000 readings
- Method Ratio Nephelometric Method (90°), ratio of scattered and transmitted light; Adaptation of the USEPA Method 180.1 and Standard Method 2130 B
- Measuring Mode normal, average, continuous
- Turbidity Standards <0 .1="" 100="" 15="" 750="" and="" li="" ntu=""> 0>
- Calibration two, three or four-point calibration
- Log Memory 200 records
- PC Connectivity USB or RS232
- Environment up to 50°C (122°F); RH max 95% non-condensing
- Power Supply 1.5V AA alkaline batteries (4) or AC adapter; auto-off after 15 minutes of non-use
- Dimensions / Weight 224 x 87 x 77 mm (8.8 x 3.4 x 3.0”) / 512 g (18 oz.)